GENOSYL (nitric oxide) is administered by inhalation. Nitric oxide is a pulmonary vasodilator. Nitric oxide is generated from liquid dinitrogen tetroxide (N 2 O 4 ) by the Cassette in the GENOSYL Delivery System. Upon initiation of GENOSYL Delivery System, the liquid N 2 O 4 is heated and the equilibrium shifts to nitrogen dioxide (NO 2 ) gas. The NO 2 is then converted into nitric oxide (NO) using the antioxidant Cartridges, and nitric oxide is delivered to the patient by means of a ventilator or a nasal cannula. The amount of nitric oxide administered to the patient is set by controlling the temperature of the N 2 O 4 liquid module, which controls the pressure inside the liquid module, which in turn controls the mass of NO 2 that is sent to the primary Cartridges, and hence the mass of nitric oxide. The mass flow of nitric oxide, together with the air from the pump, control the nitric oxide concentration. A nitric oxide sensor monitors the nitric oxide in the patient line. GENOSYL Delivery System is designed to deliver a controlled level of nitric oxide blended with breathing air or oxygen-enriched breathing air. The GENOSYL Delivery System controls the flow of nitric oxide mixed with air delivered to the patient. The structural formula of nitric oxide (NO) is shown below: Chemical Structure